VPM-4 Specifications

VPM-4 SMARTPIRANI™ TRANSDUCER SPECIFICATIONS

HOME > PRODUCTS > VACUUM > VPM-4 > VPM-4  SPECIFICATIONS

Specifications


Measuring range

1×10-6 to 13.3 mbar (7.5×10-7 to 10 Torr)

Measuring principle 1x10-6 to 10 mbar

MEMS Pirani thermal conductivity (Heat-loss)

Accuracy 1×10-5 to 9.99×10-5

25 % of reading

Accuracy 1×10-4 to  9.99×10-1 mbar

5 % of reading

Accuracy 1.00 to 13.3 mbar

25 % of reading

Maximum measuring resolution

1.0E-7 mbar

Output signal 1

1 VDC / decade (0.5VDC at 1.0E-6 mbar)

Output signal 2 (MKS 901P, 910, 925 Emulation)

1 VDC / decade (1.0VDC at 1.0E-5 Torr)

Output signal 3 (MKS GP 275 Convectron®)

0.375 to 5.659 VDC

Output signal 4 (MKS 523)

0.5 VDC / decade (MKS 523)

Output signal 5 (Inficon PSG500, PSG502, PSG510, PSG512 )

1.9 - 10 VDC (Inficon PSG55x, Leybold TTR91)

Output signal 6 (Leybold TTR91, TTR96, TTR101)

1.5 - 8.5 VDC (Pfeiffer TPR260/27x/28x)

Output signal 7 (MKS GP 275 Convectron®)

1.9 - 9.1 VDC

Output signal 8 (MKS GP 390 MicroIon)

0.5 VDC / decade

Output Signal 9 (MKS 971, 972, 974B)

0.5 VDC / decade (MKS 971B, 972B, 974B)

Output signal 10 (Inficon Gemini output 1)

1 VDC / decade

Output signal 11 (Inficon Gemini output 2)

1.33  VDC / decade

Output signal 12 (Inficon Gemini output 3)

0.6 VDC / decade

Output signal resolution

16 bit

Solid state relay contact rating (optional)

250 mA, 50 VDC / VAC peak

Environment conditions


Operating ambient temperature

-40 to +50 °C (-40 to +122 °F)

Media temperature

-20 to +50 °C (-58 to 122 °F)

Storage ambient temperature

-40 to +125 °C  (-40 to +257 °F)

Maximum media pressure

10 bar

Mounting position

Arbitrary

Protection rating, EN 60529/A2:2013

IP40

Humidity, IEC 68-2-38

98%, non-condensing

Power supply


Supply voltage 

12-30 VDC

Power consumption 

350 mW (Max)

Yes

Overvoltage protection

Yes

Internal fuse

100 mA (Thermal recoverable)

Materials


Vacuum process flange

SS 1.4307 / AISI 304L stainless steel

Enclosure

SS 1.4307 / AISI 304L stainless steel / Aluminum

Vacuum exposed materials (media wetted)

304L Stainless steel, Kovar, glass, silicon, nickel, aluminum, SiO2, Si3N4, gold, Viton®, low out-gassing epoxy resin

Process leak tightness 

<1·10-9 mbar·l/s

Approvals


CE

EMC directive 2014/30/EU

RoHS compliance

Directive EU 2015/863

Convectron® is a registered trademark of MKS Instruments Inc.


Specifications are subject to change without further notice.